Employs FID method featuring superb response characteristics and stability. Allows highly sensitive and highly precise continuous measurement. Measurement can be performed in 7 ranges and in a range ratio of 0 to 10/30,000 ppm C, making it possible to safisfy a wide variety of applications.
Components | Min.Range | Max.Range | High–sensitivity Range |
---|---|---|---|
THC | 0 – 50 ppm C | 0 – 3% C | 0 – 10 ppm C |
Configuration | Desk top or 19" panel mount |
Measuring method | Flame Ionization Detection |
Repeatability | ±0.5% F.S. (±1.0% F.S. for high–sensitivity range) |
Drift | ±1.0% F.S. per day (±2.0% F.S. per day for high–sensitivity range) |
Response time | 10 sec. (electrical) |
Ambient conditions | Temperature: 0 to 40 °C) Humidity: 90& RH or less |
Power | 100⁄115⁄220 V AC, 50⁄60 Hz, 500VA |
Output signal | 0–1V DC, 0–10V DC, 0–16 or 4–20mA DC, isolated |
No. of range | 7 ranges |
Display | LED digital display (effective 4 digits) |
Operation gas | Combustion gas should be specified; pure H2 (flow rate: about 30 mL/min.) or 40% H2⁄He (flow rate: about 120 mL⁄min.) |
Calibration gas | Zero: N2 or Air, Span: Span gas |
Dimensions | 430(W) x 132(H) x 550(D) mm, 16.9(W) x 5.2(H) x 21.7(D) in |
Mass. | 20kg, 44 lb |
Sample gas requirements | Dust: No dus |
Mist: Completely removed | |
Pressure: More than 50 mm H2O | |
Temperature: Ambient temperature | |
Moisuture: 5°C saturation or below | |
Flow rate: 500 mL⁄min | |
Exhaust gas: Discharged at ambient pressure |
For more details please contact us here or visit HORIBA Process and Environmental