DIRECT SENSE® IAQ
Indoor Air Quality Meters
The PG–300 is a compact and lightweight stack gas analyzer that can simultaneously measure up to five separate gas components. Its capability of precise measurement proves reliability in field measurements as well as its great performance in the lab facilities. Methane (CH4) analyzer is newly joined in the PG series and a total of nine models are lined up. The approx. 20% lighter body from previous model with the rugged side guards which prevent the analyzer unit from shocks and damages; the field concious design supports whatever the measurement scene is. An optional electronic cooler unit is available for long–term measurement under tough environments such as gas turbines, boilers, and incinerators facilities. PG–300 series is also suitable for solving envioronmental problems, measures of energy conservation, study of catalysts, and control of process gas.
NOx | SO2 | CO | CO2 | O2 | CH4 | ||
---|---|---|---|---|---|---|---|
Standard models |
PG–320 | • | • | ||||
PG–325 | • | • | |||||
PG–330 | • | • | • | ||||
PG–335 | • | • | • | ||||
PG–337 | • | • | • | ||||
PG–340 | • | • | • | • | |||
PG–350 | • | • | • | • | • | ||
CH4 models |
PG–324 | • | • | ||||
PG–344 | • | • | • | • |
NOx | A–range 0–25⁄50⁄100⁄250⁄500⁄1000⁄2500ppm(standard spec.) B–range 0–50⁄100⁄250⁄500⁄1000⁄2500⁄5000ppm |
SO2 CO |
0–200⁄500⁄1000ppm |
CO2 |
A–range 0–200⁄500⁄1000⁄2000⁄5000ppm (standard spec.) B–range 0–0.5⁄1⁄5⁄10⁄15 vol% (B–range is not available for PG–350) |
O2 | 0–5⁄10⁄25 vol%(Galvanic or Zirconia method) 0–10⁄25 vol%(Paramagnetic method) |
A–range | B–range |
CH4 0–2000⁄5000ppm | 0–50⁄100vol% |
CO 0–2000⁄5000ppm | 0–10⁄50vol% |
CO2 0–5⁄10⁄20vol% | 0–25⁄50⁄75vol% |
O2 0–5⁄10⁄25vol% (Galvanic or Zirconia method) 0–10⁄25vol% (Paramagnetic method) |
0–10⁄25vol% (Paramagnetic method) |
Standard Model |
CH4 Model |
|
Repeatability | ±0.5% of full scale (NOx:≥100ppm range ⁄ CO: ≥1000ppm range) ±1.0% of full scale (Except above) |
±1.0% of full scale |
Linearity | ±2.0% of full scale | |
Drift | ±1.0% of full scale ⁄ day (For SO2 analyzer only: ±2.0% of full scale ⁄ day) | |
Response Time (T90) | Analyzers except SO2 analyzer: 45 sec. or less (From sample inlet) SO2 analyzer: 180 sec. or less (From sample inlet) |
|
Sample Gas Flow Rate | Approx. 0.5L⁄min. (Approx. 1.0L⁄min for CH4 100% range) | |
Display | Measurement (3 or 4 digit display), range flow rate, etc. | |
Output | DC 4–20 mA (non–insulated) ⁄ Ethernet | |
Warm–up Time | 30 min. ±2.0% of full scale ⁄ 2 hours | |
Data Saving | SDTM memory card ⁄ SDHCTM memory card | |
Ambient Temperature | 0°C–40°C <32°F–104°F> | |
Ambient Humidity | 85% RH or less | |
Power | AC 100V to 240V, 50⁄60 Hz | |
Power Consumption | Approx. 160VA in a steady state, maximum 220VA | |
Dimentions | (With side guards)
300(W) x 520(D) x 260(H) mm <11.8"(W) x 20.5"(D) x 10.2"(H)> (Without side guards) 260(W) x 520(D) x 260(H) mm <10.2"(W) x 20.5"(D) x 10.2"(H) |
|
Weight | Approx. 13kg ∼ 15kg (29lb ∼ 33lb) | |
Sample Gas Condition | Temperature: Less than 40°C (104°F) H2O Content: Standard or less at ambient temperature Dust: 0.1g⁄m3 or less Pressure: ±0.98 kPa |
For more details please contact us here or visit HORIBA Process and Environmental